A laboratory column is used to purify a peptide and we obtain satisfactory results under the following
Question:
A laboratory column is used to purify a peptide and we obtain satisfactory results under the following conditions: velocity 30 cm/hr., bed height 15 cm, column diameter 2.5 cm, temperature 25°C, diameter of adsorbent 75 mm, and pressure drop across the bed 3.75 kg/cm2. At constant velocity, the pressure drop is inversely proportional to the particle surface area. We want to scale-up the process by increasing the bed volume a thousand-fold and the column diameter tenfold. If the same superficial velocity is to be maintained, what will the pressure drop across the bed be?
Fantastic news! We've Found the answer you've been seeking!
Step by Step Answer:
Related Book For
Principles Of Downstream Techniques In Biological And Chemical Processes
ISBN: 9781771881401
1st Edition
Authors: Mukesh Doble
Question Posted: