486 Amedical grade pressure transducer has been developed for use in invasive blood pressuremonitoring. The output voltage

Question:

4–86 Amedical grade pressure transducer has been developed for use in invasive blood pressuremonitoring. The output voltage of the transducer is vTR = ð0:06P − 0:75ÞmV, where P is pressure in mmHg. The output resistance of the transducer is 1 kΩ. The blood pressure measurement is to be an input to an existing multisensor monitoring system. This system treats a 1-V input as a blood pressure of 20 mmHg and a 10-V input as a blood pressure of 200 mmHg. Design an OP AMPcircuit to interface the new pressure transducer with the existing monitoring system.